VNA starts with S12=S21=-50dB?
Please advise.
Hi
This is true if You disconnect VNA ports
You are measuring reflections from open ports and isolation of ports
S21 & S12 represent signals from port1 to port2 and from port2 to port1. If you do not connect them toghether you see a large attenuation (-50 dB) that is the lower limit you VNA can measure.
I hope it can help.
Mazz
no,it is nor normal,it is something wrong!
Hi there,
After calibration of a network analyzer I always do a simple check with a 3dB attenuator: should measure -6dB Sii when connected either to Port1 or to Port2, and should measure -3dB Sij and Sji when connected between Port1 and Port2. If this doesn't work, then no point going any further...
It could also be a problem with the trigger if your measurement is pulsed. If you use Agilent 8753 series, the trigger is done on the falling edge and doesn't work above 50Hz...
Mosieur_Oiso