I want to see the output spectrum of a mixer using two input sources (RF & LO) from
the PNA-X and the output will be displayed IN the same network analyzer ..
because the two synthesized sweeper are not available.
pls how can we make the connetion.?
i very appriciate your help..
N
Hello nassim1:
Normally , this function is one options in the PNA-X.
You need to buy the below option.
Option 080 Frequency offset
Recommend:
Option 224, 2 ports, add internal 2nd source, combiner and mechanical switches
Option 423, 4 ports, add internal combiner and mechanical switches
Below is the AN about the R&S mixer measurement.
Many thanks for your replying me... thank you also for the document that you sent me it's so interesting...
Unfortunately, i do not have that network analyzer, i'm using Agilent PNA-X which has no same options...
Thank you again nuccu,
Nassim
Sorry for the late response but we (at Agilent PNA/PNA-X/VNA support) do not routinely monitor this particular forum web page. Regardless, , I refer you to the following publications regarding your measuremment:
1. Agilent App Note 1408-17 PNA-X IMD
http://cp.literature.agilent.com/lit...989-7265EN.pdf
2. PNA-X Online Help for PNA-X Opt 087: IMD Measurement for Convertors http://na.tm.agilent.com/pna/help/la...Converters.htm
3. If after reviewing the above information, you still have have any follow up questions, (if you reside in the United States/Canada,) he can call the Agilent Technical Contact Center (1-800-829-4444 or 1-877-894-4414 US/Canada, respectively) for additional assistance.
In the future for questions like these, please use the Agilent instrument's web forum at https://forums.tm.agilent.com/community/