微波EDA网,见证研发工程师的成长!
首页 > 研发问答 > 微波和射频技术 > 天线设计和射频技术 > Scattering parameters of test fixtures from TRL calibration

Scattering parameters of test fixtures from TRL calibration

时间:04-09 整理:3721RD 点击:
Hi all,
my first time on this forum.
I'm trying to find the scattering parameters of two test fixtures using TRL, but I cannot get the complete scattering matrix. My test fixtures are essentially coaxial to waveguide transitions.
Following exactly TRL procedure I don't get the correct value of S12 and S21 of each test fixture, but only the correct value for S12*S21. I need the correct scattering matrix of one test fixture to de-embed it from another measurement made with it. Is there another method to obtain the complete scattering of both test fixtures?
Can you suggest some reference or hint to do this.
thanks for any help

Hi,

If your Test Fixture is symmetrical you can perform a thru measurement ent get the [S] file. Then consider that is a symmetrical TF end cut the [S] file.

Another way consists in performing thru measurement and in open conditions. You have one information more.
ΓM1 = SA21*SA12*ΓOpen
ΓM2 = SB12*SB21*ΓOpen

Copyright © 2017-2020 微波EDA网 版权所有

网站地图

Top