solt calibration procedure

时间:04-01 整理:3721RD 点击:
Hallo,

for SOLT calibration, the standard 12term error model is used. Could somebody please explain why the transmission coefficient from the Network analyzer to the DeviceUnderTest is always assumed to be 1 (without loss of generality).

For better understanding I will download a picture of the 12 term error model.

Thank you.

Hallo,

Added after 1 minutes:

Hallo,

simple answer. E-TF covers both the transmission part of the error network 1 (which was set to 1) and the error network 2 (port2).