微波EDA网,见证研发工程师的成长!
首页 > 研发问答 > 微波和射频技术 > 天线设计和射频技术 > SAR Measurement Setup for Radiometric applicator in L band

SAR Measurement Setup for Radiometric applicator in L band

时间:04-05 整理:3721RD 点击:
Hi,

I'd like to measure the SAR and I want to build the measurement setup fur this purpose. The radiometric applicator is used for medical radiometer system. For example, If I have a acrylic box filled with water (40*40*40) cm^3 as model of human tissues and I put two identical applicators on the boundaries of box and connect them to the VNA ( vector network analyzer). Can I measure the SAR using this kind of setup?

Thanks in advance
Hadi

To measure Specific Absorption Rate you need to measure the Electric Field that penetrates a particular material with a specific density and a specific conductivity.
If you know the density and the conductivity of the penetrated material, the Network Analyzer setup can give you the information about the electric field (you know the generated RF power at one port, and the received power at the other). But you have to calibrate the setup for this measurement using the available probes, and this is not an easy job without having some references.

Thanks for your reply.........I'd like to use the same applicator as reference and AUT i.e. both of them have the same SAR distributions. Is it correct in this way? The problem that I don't have E or H probes in our labs and I want to find another solution as well as these probes are expensive. DO you have some references that construct SAR measurement setup in the same way what I described?

As I mentioned, if you don't have the calibrated probes, you have to use a reference to calibrate your setup. Try to rent/borrow some calibrated probes just to calibrate your setup. I do not have more information in this regards, but I am pretty sure the net is full of SAR setup examples.

Copyright © 2017-2020 微波EDA网 版权所有

网站地图

Top